Position error E18 (FAQ)

Problem:

E18: Carousel position error! Check Rinse position.

Cause:

Value of potentiometer does not match the value stored in memory of the current rinse position.

Answer:

  • Check if the rinse position is right.
  • Set correct rinse position and perform a "Learn carousel position".
  • Check mechanical connection potentiometer.

 

Additional information:

Carousel Position error

Applicable for:

Interrliner
Starrsed RS
Starrsed TL