Pipette looks like a zebra crossing (FAQ)

Question:

Pipette looks like a zebra crossing

Hint:

ESR Error 3

Answer

Normally caused by a bad O-ring

  1. Check pipette bottom, glass must not be chipped.
    • Replace pipette.
  2. Pipette bottom dirty with dried blood.
    • Clean pipette bottom
    • Check disinfectant flow
  3. Pipette bottom not straight.
    • Replace pipette.

If this happens randomly or with each pipette, check the following.

  1. Check the condition of the fill nozzle O ring.
    • Replace the fill nozzle O ring.
  2. Fill nozzle scratched, due to incorrect washer replacement.
    • Replace fill nozzle
  3. Fill nozzle alignment to pipette incorrect.
  4. Worn out fill nozzle motor.
  5. Possible too low sample volume during aspiration

Additional information :

Air bubbles

Applicable for:

Interrliner
Starrsed RS
Starrsed TL
Starrsed Compact
for Starrsed ST

See Also

Aspiration, vacuum and liquid flow FAQ index

Very small air bubbles enter the pipette during aspiration (FAQ)

One air bubble in the pipette (FAQ)

Air bubbles entering the pipette during aspiration (FAQ)

One air bubble about 5mm under meniscus (FAQ)

Liquid flowing back to the rinse bottle or cubitainer (FAQ)

Waste pump slows down and stops after 1 minute (FAQ)

First pipette is not washed sufficiently (FAQ)

Liquid dripping from needle (FAQ)

Rinse solution is not primed through system (FAQ)

Position error E18 (FAQ)

Fill nozzle does not engage with pipette (FAQ)

Fill-nozzle not functioning (FAQ)

Container "empty" not detected on Waste container (FAQ)

Sample is not or only partially aspirated (FAQ)

Waste pump tube not functioning correctly (FAQ)

Zebra striping(FAQ)

ESR unit does not accept any ESR from LAB host computer (FAQ)

Carousel moves three position steps (FAQ)

ow to handle samples if the barcode is not read correct ?