Rinse solution is not primed through system (FAQ)

Symptom :

Rinse solution is not primed through system

Hint :

  1. Check the rinse solution level in the rinse container.
    • If the level is insufficient the message Rinse bottle is empty will displayed on the screen and the alarm sounds!
  2. Check the pick-up tube in the rinse container, it may have detached from the tube connector in the lid.
    • Loosen the cap of the rinse container in order to inspect.
  3. Check the peristaltic pump tube conditions.
    • Replace every month or after 7500 samples (for Starrsed Compact 3000 samples
  4. An incorrectly fitted peristaltic pump tube may cause the problem.
  5. One of the tubes carrying the rinse solution may be blocked or loose

Additional information :

Peristaltic pump tubes replacement (WI-162 Peristaltic pump tube replacements)

Applicable for:

Interrliner
Starrsed RS
Starrsed TL
Starrsed Compact

FAQ 30014

See Also

Aspiration, vacuum and liquid flow FAQ index

Very small air bubbles enter the pipette during aspiration (FAQ)

One air bubble in the pipette (FAQ)

Air bubbles entering the pipette during aspiration (FAQ)

One air bubble about 5mm under meniscus (FAQ)

Liquid flowing back to the rinse bottle or cubitainer (FAQ)

Waste pump slows down and stops after 1 minute (FAQ)

First pipette is not washed sufficiently (FAQ)

Liquid dripping from needle (FAQ)

Position error E18 (FAQ)

Fill nozzle does not engage with pipette (FAQ)

Fill-nozzle not functioning (FAQ)

Container "empty" not detected on Waste container (FAQ)

Pipette looks like a zebra crossing (FAQ)

Sample is not or only partially aspirated (FAQ)

Waste pump tube not functioning correctly (FAQ)

Zebra striping(FAQ)

ESR unit does not accept any ESR from LAB host computer (FAQ)

Carousel moves three position steps (FAQ)

ow to handle samples if the barcode is not read correct ?

To top of page Mechatronics home Send Feedback Print