One air bubble about 5mm under meniscus (FAQ)

Symptom :

One air bubble about 5mm under meniscus.

Hint :

  1. If just one air bubble is found about 5mm below the meniscus
    • The fill speed may be too fast. Normal speed is 2-3 sec. from pinch valve to fill stop sensor. The column should not exceed the fill stop sensor by more than 10mm.
  2. Low filling speed may cause incorrect sample/diluent mixing or an abnormal number of hazy reports.

Additional information :

Air bubbles

 

 

Applicable for:

Interrliner
Starrsed RS
Starrsed TL

FAQ 30042

See Also

Aspiration, vacuum and liquid flow FAQ index

Very small air bubbles enter the pipette during aspiration (FAQ)

One air bubble in the pipette (FAQ)

Air bubbles entering the pipette during aspiration (FAQ)

Liquid flowing back to the rinse bottle or cubitainer (FAQ)

Waste pump slows down and stops after 1 minute (FAQ)

First pipette is not washed sufficiently (FAQ)

Liquid dripping from needle (FAQ)

Rinse solution is not primed through system (FAQ)

Position error E18 (FAQ)

Fill nozzle does not engage with pipette (FAQ)

Fill-nozzle not functioning (FAQ)

Container "empty" not detected on Waste container (FAQ)

Pipette looks like a zebra crossing (FAQ)

Sample is not or only partially aspirated (FAQ)

Waste pump tube not functioning correctly (FAQ)

Zebra striping(FAQ)

ESR unit does not accept any ESR from LAB host computer (FAQ)

Carousel moves three position steps (FAQ)