If just one air bubble is found about 5mm below the meniscus
The fill speed may be too fast. Normal speed is 2-3 sec. from pinch valve to fill stop sensor. The column should not exceed the fill stop sensor by more than 10mm.
Low filling speed may cause incorrect sample/diluent mixing or an abnormal number of hazy reports.