Question:
|
Zebra striping effect
|
Hint:
|
ESR error 3
|
Symptom 1:
|
During End_of_day_wash no zebra striping is visible, but during the sampling in some pipettes the blood column looks like zebra striping.
|
Answer:
|
Check if the waste pump tube is in good order.
|
Symptom 2:
|
Zebra striping occurs in the same pipette(s)
|
Answer:
|
Dirt, e.g. dried blood.
- Clean the pipette.
- Check disinfectant flow at the rinse nozzle.
Perpendicularity and straightness of the bottom face.
Glass may be chipped.
|
Symptom 2:
|
Zebra striping randomly or with each pipette
|
Answer:
|
- Check fill nozzle O-ring.
- Check fill nozzle alignment to pipette.
- Check the nozzle arm is tight on the rear vertical shaft.
|