Air bubbles entering the pipette during aspiration (FAQ)

Symptom :

Air bubbles entering the pipette during aspiration

Hint :

Can be recognized during the aspiration as blood is falling over the trumpet of the fill nozzle.

  • Check the the vacuum.
  • Check if the bottom of the pipette is dry.
    • Readjust the rinse nozzle.
  • Check the rubber V ring at the bottom end of the pipette is engage with the rinse nozzle during the wash.
  • Check the bottom of the pipette is clean.
    • If dirty, bottom pipette need to be cleaned.
  • The surface tension of the pipette glass is to high, do a Fill and clean with the cleaning agent.

Additional information :

Air bubbles

Applicable for:

Interrliner
Starrsed RS
Starrsed TL
Starrsed Compact

FAQ 30060

Related FAQ's

Very small air bubbles enter the pipette during aspiration (FAQ)

One air bubble in the pipette (FAQ)

One air bubble about 5mm under meniscus (FAQ)

Liquid flowing back to the rinse bottle or cubitainer (FAQ)

Waste pump slows down and stops after 1 minute (FAQ)

First pipette is not washed sufficiently (FAQ)

Liquid dripping from needle (FAQ)

Rinse solution is not primed through system (FAQ)

Position error E18 (FAQ)

Fill nozzle does not engage with pipette (FAQ)

Fill-nozzle not functioning (FAQ)

Container "empty" not detected on Waste container (FAQ)

Pipette looks like a zebra crossing (FAQ)

Sample is not or only partially aspirated (FAQ)

Waste pump tube not functioning correctly (FAQ)

Zebra striping(FAQ)

ESR unit does not accept any ESR from LAB host computer (FAQ)

Carousel moves three position steps (FAQ)

ow to handle samples if the barcode is not read correct ?

To top of page Mechatronics home Send Feedback Print