Liquid flowing back to the rinse bottle or cubitainer (FAQ)

Symptom :

Liquid flowing back to the rinse or saline container

Hint :

Check the condition of the peristaltic pump tubes.
They may be wrong sized, leaking, old, or worn out.
(replace once a month or after 7500 samples, 3000 for Starrsed Compact Standalone)

Attention:The wider bore tube is for the rinse pump. The narrower bore tube is for the saline pump

Additional information :

Peristaltic pump tubes replacement (WI-162 Peristaltic pump tube replacements)