Symptom :
|
Very small air bubbles enter the pipette during aspiration
|
Hint :
|
Can be recognized in the pipette as very small zebra crossing on top of blood column
|
Action:
|
- Check the inside of the fill nozzle for contamination on the O-ring.
- Clean the fill nozzle. (WI-201)
- The O ring is losing its flexibility.
- Replace the O ring in the nozzle. (WI-201)
- Check the diluter system for air bubbles.
- If air bubbles are in the diluter system, prime diluent.
|
Additional information :
|
Air bubbles
|
Applicable for:
|
Interrliner Starrsed RS Starrsed TL Starrsed Compact
|