Container "empty" not detected on Waste container (FAQ)

Symptom :

Container "empty" not detected on Waste container

Hint :

Usually this occurs with the Waste bottle.

  1. Waste alarm "container full", but the container is empty. Waste liquid is clinging to the inside of the bottle and gives false alarm.
    • Flush out Waste bottle with hot water.
  2. The detector in the cover over the waste bottle was triggered.
    • Reset the alarm.
  3. Cover over the waste bottle is open during the operation of the instrument.
    • Close the cover and reset the alarm.

Additional information :

 

Applicable for:

Starrsed RS
Starrsed Compact

FAQ 30020