Carousel Position error (service)

E18 is the error code for Carousel position error. Starrsed was not able to position the carousel. There was a difference found in the pipette memory position table and the actual measured position of the position sensor.

A position error can occur if the Starrsed is switched OFF while the carousel is moving to the next position or when the Rinse nozzle sticks up.

Warning: It is highly recommended to wash all samples after a position error due the fact the linked pipette number (patient data) is not the same as before the position error!

Position error solutions

  1. Check in tab Settings (password XXXXXX)- Carousel control the Rinse position.
    • Select Set rinse position.
    • Change Rinse position.
    • Enter the right value of the pipette number at the rinse station.
    • Close pop-up menu.
  2. Select Set rinse position
    • Change Rinse position
    • Press Learn carousel position
    • Close pop-up menu.

Check if the position error is solved: Select Go to rinse position, enter a higher value than the actual rinse position and the carousel should move without position error.

If position error is solved:

  1. Select the Service mode Icon to leave service mode.
  2. If samples or liquid is present in any pipette go to tab Maintenance - Prime/clean - Wash all pipettes which removes the liquid from the pipettes. All sample information will be lost.

If there is still a position error:

If the position is still not correct or if position errors occur frequently and the solutions above do not solve the problem, the following procedure must be carried out:

  1. Select the Service mode Icon to leave service mode.
  2. a) If liquid or samples are present in isolated pipettes, use a combination of Go to rinse position at the service tab and Maintenance -Prime rinse solution to remove the blood samples. All sample information will be lost.
    or
    b) Go to tab Settings and select Delete pipette data. The carousel stops moving. Use the sequence mentioned above to remove the blood samples.
  3. Check mechanical connection potentiometers.

If the position error is not solved after all the mentioned actions, the positioning device needs replacement.

See Also

Sampling

Air bubbles

How to handle with samples if the barcode is not read correct ?are not read (Case)

Fill time-out error

Leaking pipettes

Temperature error