High rate of hazy sample rates (FAQ)

Symptom :

High rate number of hazy sample rates.

Hint :

Usually happening after bad cleaning and poor maintenance of the instrument.

Action:

  1. Run a Fill and clean and use very hot water (80 °C)
  2. Bacterial growth in the diluter system.

If pipettes are clean and measure sensor values are correct but still a high rate of hazy reports is reported:

The measure sensor housing could be worn out. Replace the measure sensor.

Additional information :

Hazy reports

Applicable:

Interrliner
Starrsed RS
Starrsed TL
Starrsed Compact

FAQ 30025

FAQ 30025