Symptom :
|
The ESR in the pipettes looks normal, no bubbles, but the results are reported with Bubbles on top or Too many borders found or No cells / plasma found.
|
Hint :
|
- Check the measure head spring ESRI030314 is not broken
- Check if the measure head is pushed against the pipette
- Check the glass value of the measure head on an empty pipette. The value should be around 140
- Check if the measure head is at a 90° angle
|
Additional information :
|
The software is not able to get a good detection level between the empty glass, plasma and blood values. Any dirt on the pipette is detected and interpreted as a border and translated to one of the above mentioned errors
|
Applicable for:
|
Interrliner Starrsed RS Starrsed TL Starrsed Compact
|