FAQ&KB home Search Mechatronics home Send Feedback Print

What means L_err (--- 84 75 200) in the result field?

Question:

What does L_err(--- 84 75 200) means?

Answer:

L_err(--- 84 75 200) is a warning that a limit error is triggered by one of the limit error settings and no result is reported.
Description of the error message L_err(hhh/www/ttt/ccc) :

  • L_err means it is a "limit error"
  • hhh is the 30 minutes ESR
  • www is the 60 minute ESR
  • ttt is the temperature corrected 60 minute result
  • ccc is the column height

 

Example of a limit error message:

  • L_err( 42/ 84/ 75/200) means 42 mm in the 30 minute method and temperature correction 75 with a correct column height.
  • L_err(---/ 84/ 75/200) means 84 mm in the 60 minute method and temperature correction 75 with a correct column height.

 

Additional information:

Limit error settings, ESR Unit Settings and more information in Limit Error Codes and settings (KB)

Applicable for:

Interrliner
Starrsed RS
Starrsed TL
Starrsed Compact
Starrsed ST

FAQ_19021

Related FAQ's

High rate of hazy sample rates (FAQ)

Reported result is over 140 mm

Measure head goes down only halfway (FAQ)

Poor results from the measure head sensor (FAQ)

Result is send to the LIMS, despite of a leaking pipette

0mm/hr no cells or no plasma

Most results are reported with "Bubbles on top" (FAQ)

Measure sensor position error (FAQ)

Is it possible to export results to Excel ?

What effect have Limit error settings on sending results to LIMS ?

To top of page Mechatronics home Send Feedback Print