Result is send to the LIMS, despite of a leaking pipette

Question:

Despite of a leaking pipette, the result was sent to LIMS without a warning flag, how is this possible?

Answer :

When the column height was still above the 180 mm, the results are reported and no L_error is triggered. Sometimes the leakage is only a few drops and the column height is still high enough to report the results.

Additional information :

Troubleshooting Leaking pipette

Applicable for:

Interrliner
Starrsed RS
Starrsed TL
Starrsed Compact

FAQ19001