FAQ&KB home Search Mechatronics home Send Feedback Print

Reported result is over 140 mm

Symptom :

Reported result is > 140 mm and no Error 2 is reported

Hint :

This is only possible with a very high measurement after 30 minutes. This result will be calculated to a 60-minute result and reported as a > 140 mm result.

Additional information:

Example:
If in the 30-minute method 118 mm is measured, the calculated result will be 143 mm. This will not be reported as "Error 2". E2 will only be reported when there is >140 mm measured. and not when there is >140 mm result calculated

Applicable for:

Interrliner
Starrsed RS
Starrsed ST
Starrsed TL
Starrsed Compact

FAQ_19003

Related FAQ's

High rate of hazy sample rates (FAQ)

Measure head goes down only halfway (FAQ)

Poor results from the measure head sensor (FAQ)

Result is send to the LIMS, despite of a leaking pipette

0mm/hr no cells or no plasma

Most results are reported with "Bubbles on top" (FAQ)

Measure sensor position error (FAQ)

Is it possible to export results to Excel ?

What effect have Limit error settings on sending results to LIMS ?

What means L_err (--- 84 75 200) in the result field?