Cleaning diluent system

Step 1

  1. Fill a bottle with 50 ml bleach (sodium hypochlorite) and 950 ml de-ionized water. (5% solution)
  2. Remove the suction-tube from the diluent bottle and empty the diluent bottle.
  3. Place the suction tube in chlorine solution.
  4. Use the [Prime diluent] function to fill the dispenser system with the disinfectant.
  5. After the prime sequence stops, press [Prime diluent] to fill the dispenser system with the disinfectant.
  6. Leave the disinfectant in the system for 15 minutes.

Step 2

  1. Take the diluent suction tube out of the disinfectant.
  2. Wipe the tube clean and dry with a tissue.
  3. Empty the diluent bottle and refill it with hot de-ionized water (80°C).
  4. Place diluent suction tube in the diluent bottle with hot water.
  5. Use the [Prime] function.
  6. After the prime sequence stops, press "Prime” again to fill the dispenser system with the hot water.

Step 3

  1. Empty the bottle.
  2. Clean the bottle with new hot de-ionized water (80°C)
  3. Reconnect the the diluent container.
  4. Perform another [Prime] .
  5. After the prime sequence stops press [Prime] again to fill the dispenser system with the new diluent solution.