Sensor information (diluter)

The diluter assembly has the following sensors:

Name

Type

Comments

Motor up sensor

ON/OFF Led (Rx TX)

Detect the position of the motor

Motor down sensor

ON/OFF Led (Rx TX)

Detect the position of the motor

Motor pulse disk

Pulses

Use the counts from the disk

Diluter start sensor

ON/OFF Led (Rx TX)

For blood detection in the Mixing piece

EDTA flow sensor with ball bearing

ON/OFF Led (Rx TX)

Detect a positive diluent flow

EDTA electronic flow sensor

Detect the flow of the liquid

 

Mas = Magnetic approximation switch
This is a switch which becomes active when a metal object approaches the sensor. In the sensor itself there is a LED indicator that lights up when a metal object is detected.

The sensor statuses are echoed to LED's on the printed circuit board.

Label information:

Label number

Name

01

Diluter motor & disk sensor electronics

02

Diluter start sensor

03

Diluter top sensor

04

Diluter home sensor

10

EDTA flow sensor for the version Ball bearing and the electronic version

Error reference:

Name

Display error

Diluter top sensor

In Sample mode Diluter does not start

Diluter home sensor

In Sample mode Diluter does not start

Diluter start sensor

Dilution error >1000, the diluter does not start.

Diluter motor & disk sensor electronics

In Sample mode or in Prime diluent the diluter does not start or the syringe goes very slowly

EDTA Flow sensor

E26: "Diluent Flow" sensor out of range. Check/clean this sensor !