Hardware alignment syringe (diluter)

The alignment of the syringe on the diluter:

  • Remove the syringe
  • Loosen the two screws on syringe guide part ESRI060038 which holds the bottom end of the glass of the syringe
  • Replace the syringe
  • Start prime diluent.
  • During the prime strokes: fasten the two screws of part ESRI060038

See drawing ESRI060953 Diluter drive assembly