Performing Osmoscan
Start software
Important NOTE: Each time a the Osmo LOW or Osmo HIGH bottle has been disconnected (for example to refill the bottle), preform the "PRIME ALL" step before running an Osmoscan
- Check that all tubes are connected correctly to the Osmo HIGH and Osmo LOW, water and waste bottles
- Select "" on main screen. The Osmoscan screen appears
- Place a sample tube with DI water at the needle position.
- Move the sample needle to the bottom of the tube.
- Select the "Prime all" button.
Verify that both Osmo LOW and HIGH are pumped, indicated by the passing of a bubble. Check the sample feeding through the needle after introduction of an air bubble. In case of a problem check whether all tubing and connections have a free passage and are not folded or clogged. It might be necessary to open the cover plate to massage the tubes in the pinch valves. - Remove (priming) tube from the needle position.
- Rerun "" a second time
|