Performing Oxygenscan

This manufacturer’s standard operating protocol is to be validated by each lab.

Prior to use

  • The sample should not be stored at room temperature for more than 4 hours.
  • Samples should be stored at 4-7 ⁰C for a minimum of 30 minutes.
  • The Oxy ISO reagent should be stored at room temperature for at least 24 hours (to accommodate to the level of dissolved oxygen corresponding to ambient temperature)

In this chapter

Required materials

Pre-test preparations

Cleaning procedure

Preparing a blood sample Oxygenscan

Place sample Oxygenscan test

Oxygenscan

Data storage