Manual home Send Feedback Print

Previous Topic

Next Topic

Book Contents

Book Index

Place sample Oxygenscan test

  1. Bring the Bob into the upper position.
  2. Again “blow the nose”, 2x forcefully flush the connector and the tube connecting to the bottom of the Bob with air from the (60 ml) syringe to make it dry.
  3. Connect the sample syringe to the Luer-connector
  4. “wipe the first snot”. Hold a tissue under the bob to catch the first drop (this may contain some water from the tube and is not supposed to interfere with the test).
  5. Lower the Bob into the cup and inject until the syringe is empty.
  6. In the pO2 scan screen select New, fill in the sample ID, the viscosity of the Oxy ISO and press aspirate. The cup will make 3 rotations and then stop.
  7. As soon as the cup has stopped rotating gently and continuously press the syringe until fully empty. Do not remove, leave the syringe in place.
    Press Aspirate: The aspiration of the sample should precisely set the fluid level in the cup.
    In case the aspiration from the cup does not immediately catch on after fully emptying the syringe; by hand gently rotate the cup 90 degrees and back again to the previous position. This should start the aspiration-leveling of the sample.
    Oxyscan
    On this picture the edges around the small square are still too rounded, run the aspiration procedure a second time.

    ÿØÿà
    Both pictures above show sharp edges, the aspiration procedure was correct.
    Oxyscan
    Check the leveling around the Oxygensensor spot when the cup is rotated to the left and when rotated forward.
    Picture above shows correct leveling.
    Oxyscan
    Picture above indicating incorrect leveling.

    If leveling is not correct, thoroughly clean Bob and Cup with X-Clean and rub with X-Clean wetted tissue. Rinse well with DI water and rub with DI water tissue.
  8. Close the hood
  9. Click Continue and Start now
    Note: if the sample volume is set in accordance with the cell-count it is not necessary to change the gain setting.

See Also

Performing Oxygenscan

Required materials

Pre-test preparations

Cleaning procedure

Preparing a blood sample Oxygenscan

Oxygenscan

Data storage

To top of page Mechatronics home Send Feedback Print