Manual home Send Feedback Print

Previous Topic

Next Topic

Book Contents

Book Index

Pre-test preparations

Start-up

  1. Switch on Laptop/PC and Lorrca.
  2. Open the nitrogen cylinder.
  3. Lower de Bob into de Cup (to start accommodating to a 37⁰C temperature)
  4. Start software
    Cell Aggregation

    Recommendation: Run the 'Self-check' (weekly is set as default):
  5. The self-check takes around 10 minutes.
    If the Self Check fails: the pO2 scan will be greyed out in the main menu (screenshot)
    All calibration results are stored in C:Lorrca MaxSis\SelfCheck_V01.txt
    Steps of the Self-check are:
    Display sensor batch number: 200123-001 (example)
    - Step 1: Temperature check: OK (37,0⁰C )
    - Step 2: Flow check: OK
    - Step 3: Deoxygenate check: OK (down to < 0.3 % of ambient saturation level)
    - Step 4: Reoxygenate check: OK (between 97% and 103% of ambient saturation level)
    Verify during the deoxygenation steps (2 and 3) that the N2 flow readout above the bob is at 0.5 liter/minute. Adjust if necessary.
    Flow sensor check
  6. If the Self-check is canceled, the pO2 sensor is not calibrated, pO2 scans may not be correct.
  7. In case the Self-check fails, the reason may be:
    Too cold; the instrument or the DI water may be too cold for a correct verification. If this is the case it will pass the test on the second try.
    Alignment; visually verify that the sensor spot correctly stops at right in front of the fiber-optic.
    (In Hardware check screen, Lorrca IO, set Cup velocity to 1 rps and motor mode to run home)
    Damaged; verify that the sensor spot is not damaged by rubbing off the black silicone backing.
    Expired; the sensor spot may have expired, try a new sensor spot to see whether this one passes the test.
    Run the Self-check again.
  8. Perform a 'test run' with any (old) blood sample
  9. Perform the cleaning procedure after each run.

See Also

Performing Oxygenscan

Required materials

Cleaning procedure

Preparing a blood sample Oxygenscan

Place sample Oxygenscan test

Oxygenscan

Data storage

To top of page Mechatronics home Send Feedback Print