Oxygenscan

Test is performed automatically, according to test-settings.

Default test-time consists of:

  • Acclimatization for 120 seconds (=2 minutes.
  • Deoxygenation : default during 1300 seconds (= 21 minutes 40 seconds).
  • Re-oxygenation : default during 340 seconds (=5 minutes 40 seconds)

This can be edited in the settings via the Edit button next to Scan steps under pO2 scan options.

ÿØÿà

At measurement completion, result-data is displayed on-screen, the software offers an option to print (yes/no), and the data is stored on the drive of the computer.

Always correlate measurement results with the graph before reporting.

After each Oxygenscan follow the Cleaning procedure.

After the last Oxygenscan of the day, close the nitrogen gas supply on the bottle to avoid unnecessary loss of gas by a possible small leak.

See Also

Performing Oxygenscan

Required materials

Pre-test preparations

Cleaning procedure

Preparing a blood sample Oxygenscan

Place sample Oxygenscan test

Data storage