Cleaning procedure
- Prepare a 3 ml syringe with DI water.
- Replace the sample syringe with the DI syringe.
- Press , and simultaneously empty the DI syringe.
- After cleaning remove the DI syringe.
- Dry Bob, Cup and Luer syringe-connector with soft cloth tissue.
- "Blow the nose", 2x forecefully flush the connector and the tube connecting to the bottom of the Bob with air from the (60 ml) syringe to make it dry.
ÿØÿà - Lower the Bob into the Cup to keep all at 37 Celsius
|