Cleaning procedure

  1. Prepare a 3 ml syringe with DI water.
  2. Replace the sample syringe with the DI syringe.
  3. Press Clean, and simultaneously empty the DI syringe.
  4. After cleaning remove the DI syringe.
  5. Dry Bob, Cup and Luer syringe-connector with soft cloth tissue.
  6. "Blow the nose", 2x forecefully flush the connector and the tube connecting to the bottom of the Bob with air from the (60 ml) syringe to make it dry.
    ÿØÿà
  7. Lower the Bob into the Cup to keep all at 37 Celsius