E27: "Temperature" sensor out of range. Check Settings ! If the measured room temperature is out of range, E27: "Temperature" sensor out of range. Check Settings ! appears. Select Settings - General settings - Temperature (°) The displayed value is the temperature according the build-in thermometer.
Recommended is to measure the current room temperature by holding the thermometer in the carousel and before starting normal operation. Notes: The temperature in the instrument can be set between 18-35 °C, but a temperature range of 18-25 °C is preferred. It is possible that results are not send to the LIMS due to Limit Error settings in combination with this error. It is possible results are measured with a temperature which is out of the settings value. See Limit error settings (Limit error settings, Limit error settings).
|
||||||