E2

E2: Communication error! (Board: %s (%x), Command: %x, TWSR: %x E: %d)

Communication lost after 3 retries between Computer and Starrsed.

  • Power cable not connected on the communication PCB mounted on the back panel.
  • An I2C cable not connected
  • Serial cable not connected
  • No power on one of the PCB's
  • Short circuit or fault on one of the PCB's