E183

E183: Indexer motor timeout!

The indexer didn’t reach its position within a certain time limit.

  • Fuse was triggered, see instruction
  • Faulty indexer motor.
  • Indexer mechanically blocked.
  • Broken wire(s) to indexer

If this error repeatedly happens:

Check the long rail on which the indexer runs for build-up of grease and dirt. Clean rail and apply a thin layer of fresh ball bearing grease.

To check the fuse

  1. Check resettable fuses at the back. When the fuse was triggered:
    1. Switch OFF the ESR analyser unit.
    2. Close the Starrsed software.
    3. Check for mechanical obstructions of the elevator/indexer.
    4. Reset the fuse.
    5. Switch ON the ESR analyser unit.
    6. Restart the PC software.
  2. When the error occurs again, switch OFF all units and call for service.

open_to_resetable_fuse

Resetable_fuss.jpg

See Also

E2

E9

E14

E18

E24 (EHST)

E25

E69

E185

E193

To top of page Mechatronics home Send Feedback Print