E183
E183: Indexer motor timeout!
The indexer didn’t reach its position within a certain time limit.
- Fuse was triggered, see instruction
- Faulty indexer motor.
- Indexer mechanically blocked.
- Broken wire(s) to indexer
If this error repeatedly happens:
Check the long rail on which the indexer runs for build-up of grease and dirt. Clean rail and apply a thin layer of fresh ball bearing grease.
To check the fuse
- Check resettable fuses at the back. When the fuse was triggered:
- Switch OFF the ESR analyser unit.
- Close the Starrsed software.
- Check for mechanical obstructions of the elevator/indexer.
- Reset the fuse.
- Switch ON the ESR analyser unit.
- Restart the PC software.
- When the error occurs again, switch OFF all units and call for service.
|