E193

E193: Elevator motor timeout!

The Elevator didn’t reach its position within a certain time limit.

  • Fuse was triggered.
  • Faulty lift motor (EHST040981).

Elevator mechanically blocked.

  • Broken wire(s) to lift motor.
  • Faulty lift motor driver on Elevator board (Connection motor: 404, Encoder 405).

To check the fuse

  1. Check resettable fuses at the back. When the fuse was triggered:
    1. Switch OFF the ESR analyser unit.
    2. Close the Starrsed software.
    3. Check for mechanical obstructions of the elevator/indexer.
    4. Reset the fuse.
    5. Switch ON the ESR analyser unit.
    6. Restart the PC software.
  2. When the error occurs again, switch OFF all units and call for service.