E193
E193: Elevator motor timeout!
The Elevator didn’t reach its position within a certain time limit.
- Fuse was triggered.
- Faulty lift motor (EHST040981).
Elevator mechanically blocked.
- Broken wire(s) to lift motor.
- Faulty lift motor driver on Elevator board (Connection motor: 404, Encoder 405).
To check the fuse
- Check resettable fuses at the back. When the fuse was triggered:
- Switch OFF the ESR analyser unit.
- Close the Starrsed software.
- Check for mechanical obstructions of the elevator/indexer.
- Reset the fuse.
- Switch ON the ESR analyser unit.
- Restart the PC software.
- When the error occurs again, switch OFF all units and call for service.
|