Limit error settings
If it is certain that the LIMS has been programmed to handle all these exceptions correctly, these options may be set to YES. In all other situations, the options should be set to NO to avoid that results associated with exceeded limits are transmitted to LIMS and/or printed.
The screen Limit error settings has the following options:
-
- set to YES: always transmit results with time errors to the output.
- set to NO: transmit no results to the output when the ESR time is outside the selected range.
- set to YES: always transmit results with dilution errors to the output.
- set to NO: transmit no results to the output when the dilution rate is outside the selected range (set with Diluter settings).
- set to YES: always transmit results with column height errors to the output.
- set to NO: transmit no results to the output when the column height is outside the selected range ("-20"=column height 180mm, "-30=column height 170mm) .
- set to YES: always transmit results with bubbles on top warning to the output.
- set to NO: transmit no results to the output when there is generated a warning for bubbles on top.
- set to YES: always transmit results with hazy aspect to the output as specified in the settings.
- set to NO: three options are possible, Hazy >10, Hazy <25 and Hazy >25, transmit results to the output as specified.
- set to YES: always transmit results with temperature errors to the output.
- set to NO: the settings can be set/changed between a minimum of 15 and maximum of 34 °C.
When an option is set to and this limit error occurs, results will be printed/send to the LIMS.
Note: If limit error filters in the LIMS are not corresponding with the limit error settings on the instrument it is possible that not all expected results are send to the LIMS.
|