FAQ&KB home Search Mechatronics home Send Feedback Print

Protocol data and Limit error settings (KB)

Limit error YES: means that all data are send to the LIMS (in this case a measurement with hazy >25)

Limit error NO: means that data with a limit error are not send to the LIMS. In this case it can be seen that there was too much hazy, but in case other errors are only visible in the Starrsed software as L_err 13 29.28 200.

Herewith examples of result strings explanation of two protocols with and without Limit error settings.

Example of protocol R-3500 EPU

Sample data record format (202 bytes)

Parameter

# Of chars

Limit error
YES

Limit error
NO

Comment

Text distinction code I

1

"D"

"D"

 

Text distinction code II

1

"1"

"1"

 

Sample distinction code

1

"U"

"U"

 

Measurement date

8

05092018

05092018

Format = ddmmyyyy

Measurement time

6

084626

084626

Format = hhmmss

Rack no

6

001234

001234

Rack number = 1234

Tube position no.

2

05

05

Tube position in rack = 5

Inquiry mode

1

1

1

Barcode from barcode label

Sample ID number**

20

 

 

Patient number

Reserved

10

 

 

n.a.

IP Messages

6

 

 

n.a.

ESR 60 MINUTES

6

00029

00000

ESR value = 29*

HAZY CODE

5

00003

00003

HAZY" code = 3 (Hazy >25mm)

ERROR CODE

5

00007

00007

ESR error code

TEMPERATURE

5

00020

0020

Temperature in degr. Celsius = 20 degr.

SEDIMENTATION TIME

5

00030

0030

Sedimentation time in minutes = 30 min.

DILUTION RATE

5

00101

00101

Dilution rate = 101

ESR 30 MINUTES

5

00013

00000

ESR value = 013*

PIPETTE NUMBER*

5

00075

00075

Pipette number = 75

Reserved

5

 

 

n.a.

Aspiration date

8

 

 

Format = ddmmyyyy

Aspiration time

6

00..

00..

Format = hhmmss

Reserved

80

00..

00..

n.a.

* With or without temperature correction (depending on settings)

Example of protocol Mechatronics-01

Sample data record format (256 bytes)

Parameter

# of char.

Limit error YES

Limit error NO

Comment

Start of text

1

[STX]

[STX]

Start of TeXt

Text distinction code

8

ESRRE

ESRRE

 

Instrument ID

20

20x space

20x space

 

Sample ID

40

40x space

40x space

Patient number

Reserved (spaces)

15

15x space

15x space

 

Aspiration date

10

25012019

25012019

25th january 2019

Aspiration time

5

1415

1415

14:15h

E.S.R. 30 minutes (mm/½h)

5

00013

00000

30 minute ESR value

E.S.R. in mm. 60 minute (mm/h)

5

00029

00000

60 minutes ESR value

E.S.R. 60 minutes temperature corrected (mm/h)

5

00028

00000

60 minutes ESR value with temperature correction

E.S.R. 120 minutes (mm/h)

5

(5x space)

(5x space)

 

Reserved (spaces)

10

(10x space)

(10x space)

 

Sample code

5

0

0

Sample codes=0: Patient sample

Aspect code

5

3

3

Aspect code=3: Hazy>25mm

Manually added code

5

(5x space)

(5x space)

 

Pipette number

5

00075

00075

Pipette number=75

Sedimentation time (minutes)

5

00030

00030

Sedimentation time=60 minutes

Temperature

5

00020

00020

Temperature in °C = 20 °C

Dilution rate (%)

5

00101

00101

Dilution rate=101%

Column height (mm)

5

197

197

Column height = 197mm

Error code

5

7

7

ESR error code=7 (Limit error)

Limit error message (results)

30

L_err(13 29 28 197)

(30x space)

 

Reserved (spaces)

50

(50x space)

(50x space)

 

Checksum

1

S

4

Checksum calculation

End of text

1

[ETX]

[ETX]

End of TeXt

 

Total 256

 

 

 

With the Mechatronics protocol it is visible in the LIMS (in the Limit error message field) why no result is given

See Also

Reporting (KB)

ESR Error messages

ESR Error and Warning code messages

Limit Error Codes and settings (KB)

Limit error settings (KB)

Hazy reports

Analyser "HAZY" code messages

Sample codes

ESR error codes

Protocols

Protocols

Protocols

To top of page Mechatronics home Send Feedback Print