Protocol data and Limit error settings (KB)
Limit error YES: means that all data are send to the LIMS (in this case a measurement with hazy >25)
Limit error NO: means that data with a limit error are not send to the LIMS. In this case it can be seen that there was too much hazy, but in case other errors are only visible in the Starrsed software as L_err 13 29.28 200.
Herewith examples of result strings explanation of two protocols with and without Limit error settings.
Example of protocol R-3500 EPU
Sample data record format (202 bytes)
|
Parameter
|
# Of chars
|
Limit error YES
|
Limit error NO
|
Comment
|
Text distinction code I
|
1
|
"D"
|
"D"
|
|
Text distinction code II
|
1
|
"1"
|
"1"
|
|
Sample distinction code
|
1
|
"U"
|
"U"
|
|
Measurement date
|
8
|
05092018
|
05092018
|
Format = ddmmyyyy
|
Measurement time
|
6
|
084626
|
084626
|
Format = hhmmss
|
Rack no
|
6
|
001234
|
001234
|
Rack number = 1234
|
Tube position no.
|
2
|
05
|
05
|
Tube position in rack = 5
|
Inquiry mode
|
1
|
1
|
1
|
Barcode from barcode label
|
Sample ID number**
|
20
|
|
|
Patient number
|
Reserved
|
10
|
|
|
n.a.
|
IP Messages
|
6
|
|
|
n.a.
|
ESR 60 MINUTES
|
6
|
00029
|
00000
|
ESR value = 29*
|
HAZY CODE
|
5
|
00003
|
00003
|
= 3 (Hazy >25mm)
|
ERROR CODE
|
5
|
00007
|
00007
|
|
TEMPERATURE
|
5
|
00020
|
0020
|
Temperature in degr. Celsius = 20 degr.
|
SEDIMENTATION TIME
|
5
|
00030
|
0030
|
Sedimentation time in minutes = 30 min.
|
DILUTION RATE
|
5
|
00101
|
00101
|
Dilution rate = 101
|
ESR 30 MINUTES
|
5
|
00013
|
00000
|
ESR value = 013*
|
PIPETTE NUMBER*
|
5
|
00075
|
00075
|
Pipette number = 75
|
Reserved
|
5
|
|
|
n.a.
|
Aspiration date
|
8
|
|
|
Format = ddmmyyyy
|
Aspiration time
|
6
|
00..
|
00..
|
Format = hhmmss
|
Reserved
|
80
|
00..
|
00..
|
n.a.
|
* With or without temperature correction (depending on settings)
Example of protocol Mechatronics-01
Sample data record format (256 bytes)
|
Parameter
|
# of char.
|
Limit error YES
|
Limit error NO
|
Comment
|
Start of text
|
1
|
[STX]
|
[STX]
|
Start of TeXt
|
Text distinction code
|
8
|
ESRRE
|
ESRRE
|
|
Instrument ID
|
20
|
20x space
|
20x space
|
|
Sample ID
|
40
|
40x space
|
40x space
|
Patient number
|
Reserved (spaces)
|
15
|
15x space
|
15x space
|
|
Aspiration date
|
10
|
25012019
|
25012019
|
25th january 2019
|
Aspiration time
|
5
|
1415
|
1415
|
14:15h
|
E.S.R. 30 minutes (mm/½h)
|
5
|
00013
|
00000
|
30 minute ESR value
|
E.S.R. in mm. 60 minute (mm/h)
|
5
|
00029
|
00000
|
60 minutes ESR value
|
E.S.R. 60 minutes temperature corrected (mm/h)
|
5
|
00028
|
00000
|
60 minutes ESR value with temperature correction
|
E.S.R. 120 minutes (mm/h)
|
5
|
(5x space)
|
(5x space)
|
|
Reserved (spaces)
|
10
|
(10x space)
|
(10x space)
|
|
Sample code
|
5
|
0
|
0
|
=0: Patient sample
|
Aspect code
|
5
|
3
|
3
|
=3: Hazy>25mm
|
Manually added code
|
5
|
(5x space)
|
(5x space)
|
|
Pipette number
|
5
|
00075
|
00075
|
Pipette number=75
|
Sedimentation time (minutes)
|
5
|
00030
|
00030
|
Sedimentation time=60 minutes
|
Temperature
|
5
|
00020
|
00020
|
Temperature in °C = 20 °C
|
Dilution rate (%)
|
5
|
00101
|
00101
|
Dilution rate=101%
|
Column height (mm)
|
5
|
197
|
197
|
Column height = 197mm
|
Error code
|
5
|
7
|
7
|
=7 (Limit error)
|
Limit error message (results)
|
30
|
L_err(13 29 28 197)
|
(30x space)
|
|
Reserved (spaces)
|
50
|
(50x space)
|
(50x space)
|
|
Checksum
|
1
|
S
|
4
|
Checksum calculation
|
End of text
|
1
|
[ETX]
|
[ETX]
|
End of TeXt
|
|
Total 256
|
|
|
|