Limit error settings (KB)

Results with errors are normally not send to the LIMS, only an error message is given. If it is certain that the LIMS has been programmed to handle all these exceptions correctly, these options may be set to YES.
In all other situations, the options should be set to NO to avoid that results associated with exceeded limits are transmitted to LIMS and/or printed.

  • Send results when time exceeded
    • set to YES: always transmit results with time errors to the output.
    • set to NO: transmit no results to the output when the ESR time is outside the selected range.
  • Send results with dilution errors
    • set to YES: always transmit results with dilution errors to the output.
    • set to NO: transmit no results to the output when the dilution rate is outside the selected range (set with Diluter settings).
  • Send results with column height errors
    • set to YES: always transmit results with column height errors to the output.
    • set to NO: transmit no results to the output when the column height is outside the selected range ("-20"=column height 180mm, "-30=column height 170mm) .
  • Send results with bubbles on top warning
    • set to YES: always transmit results with bubbles on top warning to the output.
    • set to NO: transmit no results to the output when there is generated a warning for bubbles on top.
  • Send results with hazy aspect
    • set to YES: always transmit results with hazy aspect to the output as specified in the settings.
    • set to NO: three options are possible, Hazy >10, Hazy <25 and Hazy >25, transmit results to the output as specified.
  • Send results with temperature exceeded
    • set to YES: always transmit results with temperature errors to the output.
    • set to NO: the settings can be set/changed between a minimum of 15 and maximum of 34 °C.

When an option is set to YES and this limit error occurs, results will be printed/send to the LIMS.

Limit error YES: means that all results are send to the LIMS.

screen history

Limit error NO: means that results with a limit error are not send to the LIMS. Patient data is send with ESR Error message 1,2,3 or 7. Only with the Mechatronics protocol it is visible in the LIMS (in the Limit error message field as L_err_xxx) why no result is given

screen history

It is possible to customize the Limit error settings to be able to send results with selected errors e.g. only results with "Bubbles on top warning" and result with Hazy aspect <10:

screen history

See Also

Reporting (KB)

ESR Error messages

ESR Error and Warning code messages

Protocol data and Limit error settings (KB)

Limit Error Codes and settings (KB)

Hazy reports

Analyser "HAZY" code messages

Sample codes

ESR error codes

Protocols

Protocols

Protocols

To top of page Mechatronics home Send Feedback Print