Troubleshooting (Vacuum and liquid)

For general trouble shooting, see Vacuum errror

Check in case of E9 Air flow failure:

  • The connection between the flow sensor output port and tube
  • The three tube connections between the black T-connectors
  • The tube connections between the blue filter and the connectors
  • The tube connection from the blue filter to the fill block
  • The fill block washer
  • The fill nozzle itself
  • The tube connection from the fill nozzle and the sample tube
  • The sample tube at the pinch valve.
  • Fill nozzle o-ring
  • Pipette v-ring or pipette itself

In case of vacuum errors:

Check the fill nozzle on the O ring.
Check if the nozzle is assembled correctly.
If needed: Perform flow test and check all 84 numbers.
When this is OK and there is still no aspiration, in can be caused by a blockage in the mixing-piece.
Check in case of vacuum errors:

  • The O-ring in the fill nozzle
  • If the nozzle is assembled correctly.

If needed:
Perform flow test and check on all 84 numbers.

When this is OK and there is still no aspiration or are still vacuum errors: check other possible causes:

  • blockage in the mixing-piece.
  • blockage in the fill block washer, clean fill block
  • blockage of the vacuum line from flow sensor to the fill block
  • blockage of the blue filter, replace filter
  • blocked or wrong assembled liquid separator, clean liquid separator
  • flow sensor malfunction, replace flow sensor
  • blockage in the manifold or manifold lines, check for clogging
  • defective vacuum pump, replace pump

Vacuum stabilization errors are usually caused by a leaking system:

  • Check sample tube is installed correctly in the sample control pinch valve
  • Check the flat washer on the fill block is not damaged
  • Check sample tube is not leaking, check at first the fill nozzle connection
  • Check the sample control pinch valve is fixed on the steel plate. An O-ring is needed between the steel plate and the body of the solenoid.
  • Check the fill nozzle itself is not leaking, see WI 203

Other vacuum error symptoms:

Adjust_height_by_octagon_blocks

In This Chapter

Liquid sensor detection errors

Cleaning Liquid sensor

Vacuum

Flushing liquids