Flushing liquids

After each sample aspiration the entire system is washed automatically.
If there is no liquid flow:

  • Check that the peristaltic pumps are running. If the pump tubes are worn or leaking, replace the tubes.
  • Check that the pump tubes are installed correctly.
  • Check the tubes between the containers and pumps/valves.
  • Unscrew the cap from the container. Check the pick-up tubes in the container and that there is enough liquid in the container.
  • Check the tubes for blockages or kinks.

In This Section

Disinfectant flow

See Also

Troubleshooting (Vacuum and liquid)

Liquid sensor detection errors

Cleaning Liquid sensor

Vacuum