Rinse solution not primed through the system

Rinse pump with tubing

  1. Check the rinse tube condition. It may be worn or leaking or incorrectly fitted.
    Check the pick-up tube in the rinse container, it may have become detached from the tube connector in the cap. See WI-162 rinse tube replacement (WI-162 Peristaltic pump tube replacements).
  2. Check rinse solution level in rinse solution container.
    • If the level is insufficient, a message is displayed and the alarm sounds!
  3. One of the tubes carrying the rinse solution may be blocked or kinked.

See Also

Peristaltic pumps

ESRI099200 Peristaltic pump assembly

Rinse solution spilling over the instrument

Rinse and Saline pump failure

Sample probe is not washed after aspiration

Saline dripping from the needle assembly