Rinse and Saline pump failure

Rinse pump:

  1. Liquid flows back into the rinse container.
    • Replace the rinse pump tube.
  2. At the start of the rinse sequence the first pipette is not washed.

Saline pump:

  1. Liquid flowing back to the saline container.
    • Replace the saline pump tube.
  2. Needle is not washed sufficiently.

Note:
The wider bore tube is for the rinse pump.
The narrower bore tube is for the saline pump.

See Also

Peristaltic pumps

ESRI099200 Peristaltic pump assembly

Rinse solution not primed through the system

Rinse solution spilling over the instrument