Sample probe is not washed after aspiration
- Check saline level in saline container.
- If the level is insufficient, usually a message will be on the display and the alarm sounds!
- Check pick-up tube in saline container.
- It may have become detached from the tube connector.
- Loosen the cap of the saline container in order to inspect.
- Check the saline peristaltic pump tube condition.
- It may be worn or leak. (See Level 4 maintenance).
- An incorrectly fitted saline peristaltic pump tube may cause the same problem.
- One of the tubes carrying saline may be blocked or bent.
|