Sample probe is not washed after aspiration

  1. Check saline level in saline container.
    • If the level is insufficient, usually a message will be on the display and the alarm sounds!
  2. Check pick-up tube in saline container.
    • It may have become detached from the tube connector.
    • Loosen the cap of the saline container in order to inspect.
  3. Check the saline peristaltic pump tube condition.
    • It may be worn or leak. (See Level 4 maintenance).
    • An incorrectly fitted saline peristaltic pump tube may cause the same problem.
    • One of the tubes carrying saline may be blocked or bent.

See Also

Peristaltic pumps

ESRI099200 Peristaltic pump assembly

Rinse solution not primed through the system

Rinse solution spilling over the instrument

Rinse and Saline pump failure

Saline dripping from the needle assembly