End of day wash

When the program is shut down a pop-up screen appears with <End of day wash> message. To perform an End of day wash:

end_of_day_wash

For instruments without Osmoscan:

Follow the software instructions, the instrument is cleaned with de-ionized water

For instruments with Osmoscan:

  1. Remove both tubes on the PVP bottles.
  2. Mount the PVP tubes to the cleaning tube LORC070905 from the cleaning solution bottle.

    LORC070905 tubing
  3. Place a tube with cleaning solution at sample position.
  4. Follow the software instructions.
  5. Remove after finishing of the End-of-day wash the cleaning tube from the PVP tubes.
  6. Remove the PVP-bottles and store them at cool conditions.
    Program is closed now
  7. Empty waste bottle.
  8. Clean and dry bob and cup, drip plate and sample unit plate. Press OK