Gradient pump tube replacement
After 500 measurements/ half year.
|
|
|
Disconnect water and PVP tubing and pull up the sample needle
|
Lower the BOB and perform two times a function
[ LoRRca IO ] [ Prime all ] [ Start ]
|
|
Open the pump drawer and pull the drawer out.
|
|
|
Release the tube cassette.
|
Replace the tube assembly with new tube assembly following Install LORC070901 tubing (see also Appendix - Parts list spare- and replacement parts)
|
|