Gradient pump tube replacement

After 500 measurements/ half year.

070901 Disconnect tubing

Disconnect water and PVP tubing and pull up the sample needle

Lower the BOB and perform two times a Prime all function

LoRRCA Screen Icon Hardware check [ LoRRca IO ] [ Prime all ] [ Start ]

Open drawer

Open the pump drawer and pull the drawer out.

release_tube_cassette

replace_tube_cassette

Release the tube cassette.

Replace the tube assembly with new tube assembly following Install LORC070901 tubing (see also Appendix - Parts list spare- and replacement parts)

See Also

Tubing replacements

Install LORC070901 tubing

Install LORC070902 tubing set 1

Install LORC070903 tubing set 2

Install LORC070904 tubing set 3

After replacement of all tubing