After replacement of all tubing

  • Close the drawer.
  • Place the sample needle in sample tube.
  • Perform two times a Prime all function.

    LoRRCA Screen Icon Hardware check [ LoRRca IO ] [ Prime all ] [ Start ]

  • Check the pump function and exclude leakage, check the tubes are filled correctly without air bubbles.

See Also

Tubing replacements

Gradient pump tube replacement

Install LORC070901 tubing

Install LORC070902 tubing set 1

Install LORC070903 tubing set 2

Install LORC070904 tubing set 3