Position motor
This screen has the following sections: Select the required rack adapter what is used on the instrument.
See for rack example information Appendix - Rack adapters Make your selection, press Save settings. The following steps are needed for testing the position motor.
Go to position: Service purpose only. This is to move from the home position to any distance in millimeters from the home position. The following steps are needed for checking the barcode. Take a sample tube with barcode and a rack.
After pressing Read barcode the barcode is shown above the button. If barcode label is not read correctly see section Troubleshooting. Switch Check tube detection sensor: Use this switch to check the correct adjustment of the optical sensor that checks for the presence of sample tubes in the rack. See “Troubleshooting/ Tube sensor does not detect tubes” for further instructions. Switch Always read barcode: Use this switch to read the barcode at every position in the rack. This tube is correctly logged in the rack, even when the tube sensor is not detecting a tube. If the tube sensor is not detecting a tube and a barcode could also not be read, the tube position is logged as empty. See Service manual for further instructions This section is used to check the needle sting position through the top bracket. The following steps are needed for checking the needle sting location.
If the needle is not passing the hole see section Troubleshooting.
|
|||||||