Indexer screen

Advanced Serial output setting Service screen To Maintenance screen General  and Barcode To History screen To Sample screen To Setting screen Indexer settings Step control keys AutoCompact settings Elevator and Indexer Manual control Display error history To Reagents status screen Load and Save settings Move Indexer to rack barcode position Elevator settings Indexer rack in position setting Indexer rack test position setting Indexer rack mix position setting Indexer rack barcode position setting Indexer rack fill position setting Reset the indexer Indexer Clamp open Indexer Clamp close Move Indexer to rack fill position Move Indexer to rack mix position Home position Move Indexer to rack in position Move Indexer to rack test position close program

This menu has all the functions related to the following items.

  1. General screen (General / Barcode screen, General / Barcode screen)
  2. Elevator screen (Elevator screen, Elevator screen)
  3. Indexer screen (Indexer screen, Indexer screen)

The following Indexer settings can be set:

  1. Rack In position
    • Moves the Indexer to the rack in position.
    • The rack IN position is set at 30700, no need to change the value.
  2. Rack Test position
    • Moves the Indexer to the rack test position.
    • The rack Test position has a default value, this value can not be changed.
  3. Rack Barcode position
    • Moves the Indexer to the barcode reader position.
    • The rack Barcode position is set at 27900, no need to change the value.
  4. Rack Fill position
    • Moves the Indexer to the rack fill position.
    • The rack Fill position has a default value, this value can not be changed.
  5. Rack Mix position
    • Moves the Indexer to the mix position.
    • The rack Mix position is set at 500, no need to change the value.
  6. Rack Home
    • Moves the Indexer to the home position.
  7. Indexer Clamp
    • Open the Indexer clamp.
    • Close the indexer clamp.
  8. Actual position shows the position of the Indexer.
    • Step LEFT: The Indexer will move in steps to the left.
    • Step RIGHT: The Indexer will move in steps to the right.

Note:
When the settings need to be changed enter type a new number in the input window and test the new position again by pressing the associated button for that position.
Press Save settings to store the new values to disk and to flash Eprom when everything is correct.