Manual home Send Feedback Print

Previous Topic

Next Topic

Book Contents

Book Index

Explain Indexer screen

The following Indexer settings can be set:

  1. Rack In position
    • Moves the Indexer to the rack in position.
    • The rack IN position is set at 30700, no need to change the value.
  2. Rack Test position
    • Moves the Indexer to the rack test position.
    • The rack Test position has a default value, this value can not be changed.
  3. Rack Barcode position
    • Moves the Indexer to the barcode reader position.
    • The rack Barcode position is set at 27900, no need to change the value.
  4. Rack Fill position
    • Moves the Indexer to the rack fill position.
    • The rack Fill position has a default value, this value can not be changed.
  5. Rack Mix position
    • Moves the Indexer to the mix position.
    • The rack Mix position is set at 500, no need to change the value.
  6. Rack Home
    • Moves the Indexer to the home position.
  7. Indexer Clamp
    • Open the Indexer clamp.
    • Close the indexer clamp.
  8. Actual position shows the position of the Indexer.
    • Step LEFT: The Indexer will move in steps to the left.
    • Step RIGHT: The Indexer will move in steps to the right.

Note:
When the settings need to be changed enter type a new number in the input window and test the new position again by pressing the associated button for that position.
Press Save settings to store the new values to disk and to flash Eprom when everything is correct. Return to Service Indexer screen (Indexer screen, Indexer screen)

See Also

Indexer screen

To top of page Mechatronics home Send Feedback Print