Explain elevator screen

The following Elevator settings can be made:

  1. Elevator to Top position.
    • This function send the elevator to the top position.
    • The top position is set default at 75500, no need to change the default value.
  2. Elevator to Fill position.
    • This function send the elevator to the fill position.
    • The fill position is set default at 64000, no need to change the default value.
  3. Elevator to Home position.
    • This function send the elevator to the Home position.
  4. Elevator Reset.
    • This function reset the elevator to Home position.
  5. Mixer assembly test.
    • Switch ON and OFF the mixer motor.
  6. Tube unit test.
    • The Forward function send the tube unit to the front.
    • The Backward function send the tube unit to the back.
  7. Sample unit test.
    • The IN function move the sample arm from the home position into the needle assembly of the ESR analyser unit.
    • The OUT function move the sample arm from the needle assembly of the ESR analyser unit to the Home position.
  8. Actual position shows the position of the Elevator.
    • Step LEFT: The Elevator will move in steps to the left.
    • Step RIGHT: The Elevator will move in steps to the right.