Limit error settings

Limit error setting To Sample screen To History screen To reagent screen To Maintenance screen To General settings screen To General settings screen To Display settings To Carousel settings To Diluter settings To Limit error settings To QC settings Explanation limit errors When the selections are final, press the Save settings. To Service screen Close program

If it is certain that the LIMS has been programmed to handle all these exceptions correctly, these options may be set to YES.
In all other situations, the options should be set to NO to avoid that results associated with exceeded limits are transmitted to LIMS and/or printed.

The screen Limit error settings has the following options:

  • Send results when time exceeded
    • set to YES: always transmit results with time errors to the output.
    • set to NO: transmit no results to the output when the ESR time is outside the selected range.
  • Send results with dilution errors
    • set to YES: always transmit results with dilution errors to the output.
    • set to NO: transmit no results to the output when the dilution rate is outside the selected range (set with Diluter settings).
  • Send results with column height errors
    • set to YES: always transmit results with column height errors to the output.
    • set to NO: transmit no results to the output when the column height is outside the selected range ("-20"=column height 180mm, "-30=column height 170mm) .
  • Send results with bubbles on top warning
    • set to YES: always transmit results with bubbles on top warning to the output.
    • set to NO: transmit no results to the output when there is generated a warning for bubbles on top.
  • Send results with hazy aspect
    • set to YES: always transmit results with hazy aspect to the output as specified in the settings.
    • set to NO: three options are possible, Hazy >10, Hazy <25 and Hazy >25, transmit results to the output as specified.
  • Send results with temperature exceeded
    • set to YES: always transmit results with temperature errors to the output.
    • set to NO: the settings can be set/changed between a minimum of 15 and maximum of 34 °C.

When an option is set to YES and this limit error occurs, results will be printed/send to the LIMS.

Note: If limit error filters in the LIMS are not corresponding with the limit error settings on the instrument it is possible that not all expected results are send to the LIMS.