Manual home Send Feedback Print

Previous Topic

Next Topic

Book Contents

Book Index

Manual control

Needle control Valves and Actuators All pump switch to off Rinse pump switch Pumps Saline pump switch Waste pump switch To Maintenance screen Serial output setting Rack unit setting To Reagents status screen To Setting screen Advanced ESR Unit settings To History screen To Sample screen Service screen close program Manual control Display error history Vacuum pump switch Load settings, Save settings as and Save settings

  • Load settings is used for reloading the stored software settings.
  • Save settings as is used for storing software settings to a file. There are no restrictions for the file name.
  • Save settings is used for storing software settings after settings are changed or altered.

  1. Vacuum pump:
    ON the main vacuum pump is switched on.
    OFF the main vacuum pump is switched off.
  2. Waste pump:
    ON the waste pump is switched on.
    OFF the waste pump is switched off.
    The waste pump is used for emptying the liquid separator. Do not leave this function ON as it may cause waste pump damage.
  3. Saline pump:
    ON the saline peristaltic pump is switched on.
    OFF the saline peristaltic pump is switched off.
    Note: If vacuum pump is OFF rinse solution will spill over the Auto rack unit
  4. Rinse pump:
    ON the pipette wash peristaltic pump is switched on.
    OFF the pipette wash peristaltic pump is switched off.
    Note: If vacuum pump is OFF rinse solution will spill over the Auto rack unit
  5. All pumps OFF:
    All active pumps will be switched OFF. The waste pump is switched ON for one minute.

Return to Manual control (Manual control, Manual control) screen.

In This Section

Explain Pump control

Vacuum pump

Waste pump

Saline pump

Rinse pump

All pumps OFF

Needle control

Explain needle control

Explain Nozzle control

Valve control

Explain valves and Actuators

Valves

Actuator

See Also

Service screen

General / Barcode screen

LIMS settings

Compact settings

Display maintenance history

Display error history

Advanced

To top of page Mechatronics home Send Feedback Print