Valve control

Valve outer needle Sample control soleniod V-Valve wash section V-Valve fill section Fill actuator To History screen To Setting screen Needle control Display error history To Maintenance screen Advanced Pumps Manual control Serial output setting ESR Unit settings To Reagents status screen Valves and Actuators Service screen close program To Sample screen Elevator and Indexer Rinse actuators Load settings, Save settings as and Save settings

Valves

  1. Valve outer needle: Energizing the outer needle solenoid valve.
  2. Sample control solenoid: Sample control solenoid fill sequence energized. The function of this solenoid is to build up a vacuum in the Westergren pipette before the aspiration starts.
  3. V-Valve wash section: Pipette wash vacuum control valve, controls the main vacuum line between the wash-station and separator.
  4. V-Valve fill section: Vacuum control fill-nozzle / sample probe, controls the main vacuum line between the fill nozzle cap and separator.

Actuators

  1. Rinse actuator: Rinse solenoid active, rinse valve-block down.
  2. Fill actuator: Fill solenoid active, fill valve-block down.

Return to Valve control (Valve control, Valve control, Valve control) screen..

See Also

Manual control

Explain Pump control

Vacuum pump

Waste pump

Saline pump

Rinse pump

All pumps OFF

Needle control

Explain needle control

Explain Nozzle control

Explain valves and Actuators

Valves

Actuator