Valve control

To Sample screen To History screen To reagent screen To Maintenance screen To General settings screen To Service screen To TLAX Settings To LIMS Settings To ESR Unit Settings To Manual control To Service errors To Advanced To Pump control To Needle&nozzle control To Valves&Actuators control Explain Valve control Explain Actuator control Save and load settings

  • Load settings is used for reloading the stored software settings.
  • Save settings as is used for storing software settings to a file. There are no restrictions for the file name.
  • Save settings is used for storing software settings after settings are changed or altered.

Valves

  1. Valve outer needle: Energizing the outer needle solenoid valve.
  2. Sample control solenoid: Sample control solenoid fill sequence energized. The function of this solenoid is to build up a vacuum in the Westergren pipette before the aspiration starts.
  3. V-Valve wash section: Pipette wash vacuum control valve, controls the main vacuum line between the wash-station and separator.
  4. V-Valve fill section: Vacuum control fill-nozzle / sample probe, controls the main vacuum line between the fill nozzle cap and separator.
  5. Venting valve: Needle control valve
  6. Vacuum valve: Inner needle vacuum valve

Actuators

  1. Rinse actuator: Rinse solenoid active, rinse valve-block down.
  2. Fill actuator: Fill solenoid active, fill valve-block down.

See Also

Manual control

Pump control

help_display patient results

Vacuum pump

TLAX service settings

Waste pump

Pump B

Rinse pump

All pumps OFF

Needle Control

Explain needle control

Explain Nozzle control

Explain valves and Actuators

Valves

Actuator