TLAX service settings

General:

Calibrate

To calibrate all or one specific position of the robot arm.

Reset

All positions of the robot arm will be checked.

Robot

Manual control of the robot arm and robot gripper movement.
Choose location and press Go to position, the robot arm moves to the chosen location. With the options Pick tube, Go to Position and Release tube a sample tube can be placed manually from one position to another. This can also be used to check a calibrated position.

Track

Shows if a tube is detected on the "Track" position.

Tube disk

Shows if a tube is detected. The tube disk can be rotated manually.

Liquid sensor

Shows liquid detection in the needle valve tube.

Manual input

The manual input adapter can be moved in and out.

Gripper

To check the functioning of the gripper with "Open" and "Close".

Barcode reader:

After pressing Read barcode the barcode is shown above the button if a tube is positioned on barcode reading position (and label is in readable position)

On this screen is also presence of tubes on: Track, Tube disk, Rack input or Gripper position are shown.

See Also

Manual control

Pump control

help_display patient results

Vacuum pump

Waste pump

Pump B

Rinse pump

All pumps OFF

Needle Control

Explain needle control

Explain Nozzle control

Valve control

Explain valves and Actuators

Valves

Actuator