Problem
|
Blockage in mixing-piece
|
Hint
|
Most of the time a small blockage.
|
Action
|
Check and clean the mixing piece for small particles.
|
Problem
|
Blockage in sample tube
|
Hint
|
Most of the time a small blockage.
|
Action
|
Clean the sample tube or replace the sample tube.
|
Problem
|
Very slow aspiration of the sample
|
Hint
|
Most of the time a small blockage in the line sample probe to fill nozzle.
|
Action
|
Check the flow sensor readings. Check for small particles in the sample probe. Check for small particles in the mixing piece. Check vacuum by using a vacuum gauge.
|
Problem
|
Bad vacuum flow
|
Hint
|
Check the flow sensor readings.
|
Action
|
Check vacuum by using a vacuum gauge. Check the vacuum regulator. Check the orifices.
|
Problem
|
Sample control pinch valve error
|
Hint
|
Check the pinch valve is operational with Valve control (Valve control, Valve control, Valve control)
|
Action
|
Check the pinch valve is mounted correctly onto the stainless steel plate. Check if rubber O ring under the valve body is installed. Check the wiring of the pinch valve is not broken.
|