Diluter start sensor test

Remove the sensor from the mixing piece.

See Menu selection (diluter) to perform this task.

  • In air a low value is expected.
  • When a blockage is generated between the sensors, you expect a high value (>1000)

When the sensor is fixed on the mixing piece the value of the Diluter start sensor is 400-700.

When this test fails, replace the Diluter start sensor.

See Also

Troubleshooting (diluter)

Diluter not started

Air bubbles in diluter

Diluter errors

Diluter home sensor

Diluter motor

Diluter motor pulse disk

Diluter top sensor