Diluter start sensor testRemove the sensor from the mixing piece. See Menu selection (diluter) to perform this task.
When the sensor is fixed on the mixing piece the value of the Diluter start sensor is 400-700. When this test fails, replace the Diluter start sensor.
|
||||||